Chapter 8 - Replacement Procedures 8-175
D
IRECTION 5771498-100, REVISION 6 VENUE™ SERVICE MANUAL
PRELIMINARY
7.) Attach grommet to the ECG USB cable.
8.) Fold the excessive cable and use the cable clips to hold it.
9.) Attach the metal bracket to close the cable opening.
10.)Route the USB cable as shown below and connect the ECG USB cable to the ECG.
Figure 8-189 Attaching Grommet to the ECG USB Cable
Figure 8-190 Attaching Metal Bracket
Figure 8-191 Connecting ECG to ECG USB Cable