<Doc_TEC><Doc_502606900><Rel_B><Lang_GB>
Page 102 of 174
V1453
Electrical cabinet
Measuring
cell
Conductivity instrument
V1312
Removing a measuring cell
For easy access to the measuring cell, slacken the two screws that keep the electrical
cabinet in place. Remove the right-hand screw. Move the electrical cabinet to the left and
rotate it forward. Unscrew the measuring cell. The cables are long enough to allow the
measuringcelltobeimmersedinavesselofcalibrationliquid.