CALIBRATION
Conductivity meter
Check the output signal from conductivity meter
To check the output signal, proceed as follows.
1. Press [ on the conductivity meter.
2. Enter code 22 with Y and Z.
3. Press [. Setup 1 appears.
4. Choose Output med Y and Z.
5. Press [. Lin appears on the upper line and Sel. Type appears on the lower line.
6. Press [.
7. Choose 4-20 mA with Y and Z.
8. Press [ .
9. Set 0/4 mA = 0.000 µS/cm with Y and Z.
10. Press [.
11. Set 20 mA = 200 µS/cm with Y and Z.
12. Press [.
13. Press Y and Z at the same time to log out.
Setting the cell constant
To set the cell constant, proceed as follows:
1. Press [ .
2. Enter code 22 with Y and Z.
3. Press [. Setup 1 appears.
4. Press [ until the display shows Cellconst.
5. Set the relevant cell constant with Y and Z;seethecalibrationcerticate
6. Press [ three times. The display shows Setup 1.
7. Press Y and Z at the same time to log out.