EasyManua.ls Logo

Getinge 86 SERIES - Aborting Program Start; Aborting a Running Program

Getinge 86 SERIES
228 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
<Doc_TEC><Doc_502606900><Rel_B><Lang_GB>
Page 8 of 174
V1513
1 Pre-rinse 1
2 Pre-rinse 2
3 Wash
4 Neutralize
5 Post-rinse 1
6 Post-rinse 2
7 Final rinse
8 Disinfection
9 Drying
Program cycles
A Alkaline detergent
B Neutralisation
C Instrument milk
(extra equipment)
If instrument milk is
dosed, neutralisa-
tion is not dosed.
90
80
70
60
50
40
30
20
10
100
10 20 30 40 50 60 min
5
CW HWHW
(DW
option)
A
B
9
3 861
HWCW+HW
7
(C)
Aborting program start
A started program can be aborted within ten seconds of the door being locked. To abort
a started program, press V again. During the period when the program can be abort-
ed,ayellowlampashesat M. The door is unlocked and opened automatically and
the machine can be restarted in the usual way.
Aborting a running program
While a program is running, the machine can be stopped with the main switch.
If the power is switched on again within 60 seconds, the program continues un-
til the end. If the power is switched back on after more than 60 seconds, fault code
F00 PWR FAILUREisdisplayed.Toacknowledgethealarm,rstsilencetheaudible
signal by pressing U. When the fault has been corrected, enter the password and con-
rmbypressingS.Allliquidisdrainedfromthemachineandthesoiled-sidedooris
unlocked and goes down.