EasyManuals Logo

Getinge 86 SERIES Technical Manual

Getinge 86 SERIES
228 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #123 background imageLoading...
Page #123 background image
Page 123 of 174
<Doc_TEC><Doc_502606900><Rel_B><Lang_GB>
Measuring cell
CLS 15
Conductivity instrument
CLM 223
Removing a measuring cell
For easy access to the measuring cell, slacken the two screws that keep the electrical
cabinet in place. Remove the right-hand screw. Move the electrical cabinet to the left and
rotate it forward. Unscrew the measuring cell. The cables are long enough to allow the
measuringcelltobeimmersedinavesselofcalibrationliquid.

Table of Contents

Other manuals for Getinge 86 SERIES

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Getinge 86 SERIES and is the answer not in the manual?

Getinge 86 SERIES Specifications

General IconGeneral
BrandGetinge
Model86 SERIES
CategoryCleaning Equipment
LanguageEnglish

Related product manuals