13. Now the display shows:
A beep is heard and the display shows:
SW Update 0x2497
Updating CRC....
14.Ifloadingalanguageleaswell,gotostep8underLoadingalanguagele.
15. Check that the battery jumper (X30) is set to ON.
16. Switch off the power to the machine with the main switch.
17. Change the jumper (X29) to Cold.
18. Switch on the power to the machine with the main switch. The display should
now show:
SW Update 0x2497
CRC OK
P00
19. Without switching off the power, move jumper X29 to the Normal position.
17. Close Flashloader.
18. Start CS-1000 and load wash programs; see instructions for CS-1000.
Alterna-
tively
P01 OP-SHORT-D
60.0°C 4
New card Existing card