EasyManuals Logo

Getinge 86 SERIES Technical Manual

Getinge 86 SERIES
228 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #83 background imageLoading...
Page #83 background image
Page 83 of 174
<Doc_TEC><Doc_502606900><Rel_B><Lang_GB>
5. Enter the value 20 (C) and click Set Low.
6. Check that the value (D) rises to 20.
7. Replace resistors for 20 ºC with resistors for 90 ºC.
8. Enter 90 (E) in CS 1000 and click Set High.
9. Check that the value (F) rises to 90.
10. Log off CS 1000.
11. Remove the resistors and reinstate the temperature sensors.
V1572
V1571
C
E
F
D

Table of Contents