7. Check that the reading of the external thermometer stabilizes. Enter the reading of
the external thermometer (C) and click Set Low.
8. Check that the value (D) rises to the set value.
9. Move the temperature sensors and the external thermometer to the oil bath. Check
that the reading of the external thermometer stabilizes. Enter the reading of the ex-
ternal thermometer (E) and click Set High.
10. Check that the value (F) rises to the set value.
11. Log off CS 1000.
12. Re-instate the temperature sensors.
V1572
V1571
C
E
F
D