Maintenance
Cleaning
Always disconnect the unit before cleaning. Use only a mild solution of detergent and water applied
with a soft damp cloth. Dry thoroughly before use.
Do not use aromatic hydrocarbons or chlorinated solvents. These products may damage the unit.
Do not use alcohol or alcohol based products on the front panel, especially the display. These
products may damage the unit.
Replacing fuses
The RCS uses two T1AL 250V crystal 5x20mm fuses. The fuses are placed in the back of the
equipment, on the right side (See section Description of the Equipment Components).
Before removing the fuses case, make sure you removed the power connector (See section
Description of the Equipment Components).
Then remove the case pressing the tabs placed at the top and bottom of the case using a sharp
object.
Once removed the fuses case, pull the damaged fuses out, then put the new ones and replace the
case. Make sure the tabs are properly positioned.