GDM-8261 User Manual
116
Pin6
FAIL signal Output. Activates when the
compare result is FAIL.
Pin7
HIGH Limit FAIL signal Output. Activates
w
hen the compare result is FAIL due to
v
iolating the HIGH Limit.
Pin8
LOW Limit FAIL signal Output. Activates
w
hen the compare result is FAIL due to
v
iolating the LOW Limit.
Pin9
EOM (End Of Measurement) signal Output.
A
ctivates when compare measurement is over.
EOM pulse
w
idth timing
>10μs (R=1k)
0V
+5V
Application: Compare measurement
Applicable to
Background
Compare measurement checks and updates if the
measurement data stays between the upper (high) and
lower (low) limit specified.
1. Activate
Compare
measurement
Press the Shift key, then the Hold
(Comp) key.
HOLD
COMP
2. High limit
setting
1st display
Shows the high limit value
2nd display
Indicates high limit setting
1. Use the Left/Right key to move
the cursor (flashing point)
between high/low setting, digits,
and decimal point.