DIGITAL I/O
117
2. Change the parameter using the
Up/Down key.
3. Press the Enter key to confirm
editing and move to the low
limit setting.
3. Low limit
setting
1st display
Shows the low limit value
2nd display
Indicates low limit setting
Set the low limit in the same way as
in the high limit. Press the Enter key
to confirm editing. The compare
measurement starts right away.
4. Compare
measurement
appears
COMP
Indicates Compare mode
2nd display
Shows the compare measurement result:
Pass, High, or Low.
5. Result High
If the 2nd display shows
High, the result is above
the High limit.
Digital I/O: FAIL Out (Pin 6) and
HIGH Limit FAIL Out (Pin 7) are
activated.
Low
If the 2nd display shows
Low, the result is below
the Low limit.