Connect the sensor to a cable
l
Pin 3: Not connected
l
Pin 4: Not connected (TEDS data)
l
Pin 5: Not connected (TEDS ground)
l
Pin 6: Signal +
l
Pin 7: Not connected
l
Pin 8: Not connected
Connecting the TEDS chip:
In addition, there are two ways to connect the TEDS chip:
l
0-wire: TEDS chip is inserted between pins 2 and 7
l
1-wire: TEDS chip is inserted between pins 4 and 5
EMC-immune cabling
To have an EMC-immune cabling, a special wire assignment is used. See "Testing with high EMC immunity"
on page16 for more information.
Potentiometers (5-wire)
Board type: B201-10
Connector type: 10×RJ45 (each with 8 pins)
Potentiometers can only be connected in a 5-wire configuration, the pin out is the same as the 5-wire
configuration shown for half-bridge sensors.
Voltage transducers
Board type: B201-10
Connector type: 10×RJ45 (each with 8 pins)
Voltage transducers can be connected in an 'active' 4-wire configuration or in a 'passive' 2-wire configuration
with just the Signal + and Signal wires connected.
Active 4-wire configuration: