99
────────────────────────────────────────────────────
7.4 Command
────────────────────────────────────────────────────
Note
When the test time exceeds 999.9 s, "999.9" is returned.
If the command is sent with no decision results given, a response of <6> is returned.
*For the case that a test is stopped before a decision result is given, or others
Queries the measured voltage value.
Syntax
:MEAS:VOLT?
Response
<data>
data
Measured voltage value( Refer to section 7.5. )
Function
Queries the measured voltage value.
Example
PC
>
3158
>
:MEAS:VOLT?
1.50 Measured voltage value is 1.50 kV.
Queries the measured current value.
Syntax
:MEAS:CURR?
Response
<data>
data
Measured current value( Refer to section 7.5. )
Function
Queries the measured current value.
Example
PC
>
3158
>
:MEAS:CURR?
5.00 Measured current value is 5.00 mA.
Queries the test time elapsed.
Syntax
:MEAS:TIM?
Response
<data>
data
Test time elapsed ( Refer to section 7.5. )
Function
Queries the test time elapsed.
Example
PC
>
3158
>
:MEAS:TIM?
30.0 Test time elapsed is 30.0 s.
Note
When the test time exceeds 999.9 s, "999.9" is returned.
:MEAS:VOLT?
:MEAS:CURR?
:MEAS:TIM?