4.9 Volt age Comparator Position
────────────────────────────────────────────────────
Setting procedure
0: Not test(Initial setting)
1: End test
NOTE
4.9 Voltage Comparator Position
When performing a voltage test, you can select whether you want to view the
output-voltage when the test starts, or when the test ends.
If "0: Start test" is selected, the Voltage Comparator function is active immediately
before and during the withstand voltage test.
If "1: End test" is selected, the Voltage Comparator function is only active when the
withstand voltage test ends.
(1) Press
SHIFT
+
STOP
keys while in READY state to display the Optional
function setting screen.
(2) Use the
/ keys to move the flashing cursor to the position shown in the
figure.
(3) Use the
/ keys to set a value at the flashing cursor location.
(4) To complete the optional settings, press
SHIFT
+
STOP
keys. The unit
reverts to the READY state.
When the Voltage Comparator function is set to OFF or the Voltage Comparator
function is set to ON but the timer is set to OFF, the Voltage Comparator function
is disabled and optional settings are ineffective.