28
────────────────────────────────────────────────────
3.3 Setting the "SETTING" State
────────────────────────────────────────────────────
3.3.3 Setting the Test Time
Test Time: 60.0s
Test Time: OFF
NOTE
(1) If the flashing cursor does not appear in the READY state, display the cursor at
the position where the Test Time is indicated by pressing the
/ keys.
(2) Set the test time using the
/ keys.
With time set the time changes in 0.1 s increments (1 s increments when the set
time scale is 100 s to 999 s).
With time set at 0.5 s to 99.9 s, press
SHIFT
+ / keys. The time
changes in 1.0 s increments (10 s increments when the set time scale is 100 s to
999 s).
Settings may be made along a scale ranging from 0.5 s to 999 s (in gradations of
0.1 s for the range 0.5 s to 99.9 s and 1 s for the range 100 s to 999 s).
If no testing time is required, turn off the
ON/OFF
key.
(3) When settings are complete, press the
STOP
key.
If a test time has been set, the reduction timer will operate during the test.
If the test time is set to OFF, the time elapsed during the test is displayed. When
this time exceeds 999 s, "_" will appear, but the test will continue.
If the test time is set to OFF, the comparative-voltage value becomes ineffective.