3.3 Setting the "SETTING" State
────────────────────────────────────────────────────
3.3.3 Setting the Test Time
Test Time: 60.0s
Test Time: OFF
NOTE
(1) If the flashing cursor does not appear in the READY state, display the cursor at
the position where the Test Time is indicated by pressing the
/ keys.
(2) Set the test time using the
/ keys.
With time set the time changes in 0.1 s increments (1 s increments when the set
time scale is 100 s to 999 s).
With time set at 0.5 s to 99.9 s, press
SHIFT
+ / keys. The time
changes in 1.0 s increments (10 s increments when the set time scale is 100 s to
999 s).
Settings may be made along a scale ranging from 0.5 s to 999 s (in gradations of
0.1 s for the range 0.5 s to 99.9 s and 1 s for the range 100 s to 999 s).
If no testing time is required, turn off the
ON/OFF
key.
(3) When settings are complete, press the
STOP
key.
・
If a test time has been set, the reduction timer will operate during the test.
・
If the test time is set to OFF, the time elapsed during the test is displayed. When
this time exceeds 999 s, "_" will appear, but the test will continue.
・
If the test time is set to OFF, the comparative-voltage value becomes ineffective.