EasyManua.ls Logo

Hioki 3158 - FAIL State

Hioki 3158
136 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
40
────────────────────────────────────────────────────
3.5 PASS or FAIL Determination
────────────────────────────────────────────────────
3.5.3 "FAIL" State
WARNING
Even when a test has been terminated, there may still be voltage in
the output-voltage terminal when the DANGER lamp is lit. Before
touching the output-voltage terminal, test lead, or tested object, make
sure the analog voltmeter is at 0 kV, the DANGER lamp is OFF, and
the TEST lamp is OFF.
The FAIL state is held using the FAIL Hold function. Refer to Section "4.2 FAIL Hold Function".
If the measured voltage deviates from the upper- or lower-level value during the test,
the unit switches to the FAIL state and immediately stops outputting a voltage. The
FAIL state can be divided into UPPER FAIL and LOWER FAIL states.
UPPER FAIL indicates that the measured voltage has exceeded the upper-level
value.
LOWER FAIL indicates that the measured voltage has dropped below the lower-
level value.
(When the lower-level value is set to OFF, LOWER FAIL is ineffective.)
With a comparative voltage value set, if the output voltage deviates from the
comparative-voltage range, the unit switches to the FAIL state and UPPER,
LOWER,andFAIL light up.

Table of Contents

Related product manuals