12
────────────────────────────────────────────────────
2.6 Connecting the 9615 H.V. TEST LEAD
────────────────────────────────────────────────────
DANGER
To avoid any life-threatening electric shock accidents, ensure that the
following rules are observed.
The AC Withstanding Voltage HiTester is a dangerous product which
discharges high voltages. To prevent getting electrocuted, always
wear high-voltage protective rubber gloves when carrying out any
operation.
Be careful when using the product and ensure that you do not touch
this product, any tested object that is connected or any H.V. TEST
LEAD, etc.
WARNING
To prevent electrical shock, turn off the power unit and tested object,
make sure that there is no high voltage being applied to the output,
confirm the following 3 items, and connect the 9615 H.V. TEST LEAD.
(1) The analog voltmeter reads 0 kV.
(2) The DANGER lamp is OFF.
(3) The TEST lamp is OFF.
Before connecting the 9615 H.V. TEST LEAD, be sure to check its
insulation for tearing and metal exposure.Using the product in such
conditions could cause an electric shock, so contact your dealer or
Hioki representative for repair.
To avoid electric shock, make sure the 9615 H.V. TEST LEAD is
securely connected before starting a test, as a loose test lead can
cause a hazard when a voltage is output.
2.6 Connecting the 9615 H.V. TEST LEAD