102
────────────────────────────────────────────────────
7.6 Command Summary
────────────────────────────────────────────────────
Command Data Explanation Page
:CONF:CLOW?
Queries the lower level value
enablement.
91
:CONF:TIM
<data>
0.5 - 99.9
( Numerical data in NR2
format, two or three digits )
100 - 999
( Numerical data in NR1
format, three digits )
Test time ( Refer to section
7.5. )
Sets the test time.
92
:CONF:TIM?
Queries the test time.
92
:
STAR
Starts a test.
93
:STOP
Performs forcible ending of a test
and releases the hold state.
93
:STAT?
0: PASS
1: UPPER FAIL
2: LOWER FAIL
3: READY
4: TEST
5: VOLTAGE SETTING FAIL
6: ELSE
Queries the state.
94
:MEAS?
Queries the measured value.
94
:MEAS:VOLT?
Queries the measured voltage
value.
95
:MEAS:CURR?
Queries the measured current
value.
95
:
MEAS:TIM?
Queries the test time elapsed.
95