132
FunctionSpecications
(13) Automatic power saving
Operation Automaticallyturnsotheinstrumentifinactivity
continues or a measurement-current anomaly continues
to be detected for 10 minutes (±1 minute).
Display The [APS] segment appears.
Default setting Enabled
Conditions
where the
function is
disabled
(The function is
automatically
disabled)
During communications
While the measurement recording prompt is operating
How to set See “1.4 Names and Functions of Parts” (p. 28).
(14) Instrument’s battery level detection
Operation Displays instrument’s battery level
(The accuracy can be guaranteed until the
indicator blinks)
Display Four-level indicator (For LR6 Alkaline batteries).
10.1 V or more
9.2 V to 10.1 V
8.0 V to 9.2 V
(blinking) 7.6 V to 8.0 V
(Shutdown) Less than 7.6 V
The indicator has an error of ±0.2 V.
How to disable None
(15) Date and time
Operation 24-hour clock; leap years are adjusted automatically.
Accuracy Approx. 4 minutes per month
HIOKI BT3554F961-00HIOKI BT3554F961-00