Replacing the Tip Pin of the Test Lead
For the L2020
1
Turn o the instrument and remove the test lead.
2
Rotate the grip to loosen.
Please beware of injury since
the tip has a sharp point.
3
Remove the tip pin.
Model 9465-90 Tip Pin
4
Insert a new tip pin.
Model 9465-90 Tip Pin
5
Rotate the grip to fasten.
Fasten until it stops.
6
Check if it has been rmly xed. Loose connections could
cause a wire break or poor contact.
7
Check for proper operation.
Measure a non-defective object to check if the measured
resistance is correct before use.