XNX Universal Transmitter
Section 5 - Warnings/Faults
116
Fault Description
Appli-
cable
Sensors
Latching / Non-
Latching
Frequency of
Diagnostic
Event History Data Action For Resolution
F126
SensorPathOb-
scured
Optima Non-latching
XNXpollssensor
every2seconds,
diagnosticfre-
quencycontrolled
bysensor
Sensorfaultorwarn-
ingcode(Note4)
Checklocationforexternalinterference�Checksensorfordirtywindows�
F127 BeamBlock Excel Non-latching
XNXpollssensor
every2seconds,
diagnosticfre-
quencycontrolled
bysensor
Sensorfaultorwarn-
ingcode(Note4)
ChecklocationforexternalinterferenceorobstructionsintheIRpath�
Checksensorfordirtywindows�CheckExcelalignment�
F128
SensorInstallation
NotComplete
Excel Non-latching
XNXpollssensor
every2seconds,
diagnosticfre-
quencycontrolled
bysensor
Sensorfaultorwarn-
ingcode(Note4)
CheckExcelalignment�Performazerocalibration�
F130
OptionCommuni-
cationFailure
All Non-latching 2seconds
OptionmoduleID:
0=None,1=Foun-
dation
TM
Fieldbus,
2=Modbus
®
,3=Relay
CheckthatinstalledoptionmatchestheoptionindicatedintheXNXpart
number�Iftheoptionhasbeenchanged,thenewoptionmustbesetup
inInformation->TransmitterDataasdescribedinthemanual�
F133 Notused
F143
Stabilization
Timeout
All Latching 2seconds
Warmuptime(sec-
ondsx100)
Cyclepower,contactHoneywellAnalytics’ServiceDepartmentifprob-
lempersists�
F145 ReexFailure ECC Non-latching
Dependenton
sensor,typically8
hours;Oncefault
isdetected:every
15minutes
nA/mV ECCsensorisnolongerfunctioningproperly�Replacesensor
F146
UnknownSensor
Failure
Optima,
Excel
Non-latching 2seconds
Sensorfaultorwarn-
ingcode(Note4)
ContactHoneywellAnalytics’ServiceDepartment�
F148
Internaloption
boardhardware
failure
All Non-latching 2seconds
Optionboarderror
status(Note6)
ContactHoneywellAnalytics’ServiceDepartment�

Table of Contents

Other manuals for Honeywell XNX

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Honeywell XNX and is the answer not in the manual?

Honeywell XNX Specifications

General IconGeneral
Ingress ProtectionIP66/67
Product TypeUniversal Transmitter
Enclosure RatingIP66/67
DisplayLCD
Communication ProtocolsHART, Modbus, Foundation Fieldbus
Power Supply24 VDC nominal, 18-32 VDC
Operating Temperature-40°C to +65°C (-40°F to +149°F)
MaterialAluminum or Stainless Steel
Gas TypesToxic, Oxygen, and Combustible gases
Output4-20 mA
Relays3 programmable relays
ApprovalsATEX, IECEx, UL, CSA
Sensor CompatibilityElectrochemical, catalytic, infrared, and other sensor types