EasyManuals Logo

Honeywell XNX Technical Manual

Honeywell XNX
200 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #141 background imageLoading...
Page #141 background image
XNX Universal Transmitter
Section 6 - Specications
139
Gas type Part Number Gas Type Applied Concentration Unit Reading Unit
NO XNXXSM1SS
CarbonMonoxide 300
ppm
0 ppmNO
SulfurDioxide 5 0 ppmNO
NitrogenDioxide 5 <1.5 ppmNO
HydrogenSulfide 15 ~1.5 ppmNO
NO
2
XNXXSN1SS
CarbonMonoxide 300
ppm
0 ppmNO
2
HydrogenSulfide 15 ~-1.2 ppmNO
2
SulfurDioxide 5 0 ppmNO
2
NitrogenMonoxide 35 0 ppmNO
2
Chlorine 1 ~1 ppmNO
2
H
2
XNXXSG1SS
CarbonMonoxide 300
ppm
<60 ppmH
2
HydrogenSulfide 15 <3 ppmH
2
SulfurDioxide 5 0 ppmH
2
NitrogenMonoxide 35 »10 ppmH
2
NitrogenDioxide 5 0 ppmH
2
Chlorine 1 0 ppmH
2
HydrogenCyanide 10 »3 ppmH
2
HydrogenChloride 5 0 ppmH
2
Ethylene 100 »80 ppmH
2

Table of Contents

Other manuals for Honeywell XNX

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Honeywell XNX and is the answer not in the manual?

Honeywell XNX Specifications

General IconGeneral
Ingress ProtectionIP66/67
Product TypeUniversal Transmitter
Enclosure RatingIP66/67
DisplayLCD
Communication ProtocolsHART, Modbus, Foundation Fieldbus
Power Supply24 VDC nominal, 18-32 VDC
Operating Temperature-40°C to +65°C (-40°F to +149°F)
MaterialAluminum or Stainless Steel
Gas TypesToxic, Oxygen, and Combustible gases
Output4-20 mA
Relays3 programmable relays
ApprovalsATEX, IECEx, UL, CSA
Sensor CompatibilityElectrochemical, catalytic, infrared, and other sensor types