Chapter 2 Operating Your Universal Counter
Using the LIMITS Menu Keys
Operating Guide 2-39
2
6 Press Limit Modes key until
LIM TEST: ON
is displayed.
Note that once either the upper or lower limits have been set, the limit
testing is automatically enabled as indicated by the
LIM TEST: ON
display, and the Limit Modes indicator.
7 Press Limit Modes key.
ON FAIL: GO ON
is displayed.
8 Press any one of the arrow keys to toggle to the next state.
ON FAIL: STOP
is displayed. The
ON FAIL: STOP
function tells the
Counter to stop making measurements when the input signals drift more
than ±10% apart.
9 Press Run key to start making measurements with this limits-testing
choice selected.
Now, if a measurement drifts out of the user-entered limits the Limit
annunciator in the display will light. The Counter will stop making
measurements.

Table of Contents