Chapter 2  Operating Your Universal Counter
Preset Values After Power-Up and *RST
2-86 Operating Guide
2
Phase Arming Parameters
Phase arm yes IMMediate AUTO no
Phase ext start arm slope yes POS POS no
Pulse, Rise, Fall, Dutycycle 
Arming Parameters
Pulse Arming Parameters
Pulse start arm yes IMMediate AUTO no
Pulse ext start arm slope yes POS POS no
Time Interval Arming 
Parameters
Time Interval start arm yes IMMediate AUTO no
Time Interval ext start arm slope yes POS POS no
Time Interval start arm delay yes IMMediate NONE no
Time Interval start arm delay 
time
yes 100 E-9s 100 E-9s no
Time Interval start arm delay 
events
yes 1 1 no
Time Interval stop arm yes IMMediate AUTO no
Time Interval ext stop arm slope yes POS POS no
Time Interval stop arm delay yes IMMediate NONE no
Time Interval stop arm delay 
time
yes 100 E-9s 100 E-9s no
Time Interval stop arm delay 
events
yes 1 1 no
Input Coupling Parameters
Channel 1 input coupling yes AC AC no
Channel 2 input coupling yes AC AC no
Table 2-7B. HP 53132A (S/N 3646 and above) Preset Values (continued)
Description
In 
Save/Recall
Value at *RST 
(HP-IB Reset)
Value at 
Power-Up 
In 
Non-Volatile 
Memory