INFICON Cleaning and maintenance | 8
UL6000-Fab,-PLUS-Operating-instructions-iinc75en1-01-(2206) 117 / 152
3 Capillary filter (metal with gasket;
option)
6 Gasket
A blockage of the sniffer tip can have the following causes:
Capillary filter blocked -> replace felt discs or capillary filter
Sinter filter clogged -> check or replace sinter filter
Capillary clogged in sniffer tip -> replace sniffer tip
Sniffer tip damaged -> replace sniffer tip
Sniffer line damaged / blocked -> replace SL 200.
Replace felt discs or
capillary filters
• Switch off the leak detector or disconnect the sniffer line from the device.
• Use tweezers to remove the dirty felt discs and replace them with new ones.
or
• Unscrew the capillary filter and replace it with a new one. Do not forget the gasket
for the metal version!
Check or replace the
sinter filter
1 Switch off the leak detector or disconnect the sniffer line from the device.
2 Remove the two cross-head screws.
3 Remove the sinter filter with gasket.
4 Check the filter optically for soiling.
ð If you do not see any soiling or only slight soiling, continue to use the sinter
filter including gasket.
Fig.15: Sintered filter slightly soiled
ð With heavy soiling, replace sinter filter including gasket.
Fig.16: Sintered filter heavily soiled
5 Tighten the sniffer tip again.
ð The sniffer line is ready to be used again.