5 | Installation INFICON
40 / 152 UL6000-Fab,-PLUS-Operating-instructions-iinc75en1-01-(2206)
WARNING
Fire hazard due to overload, short circuit and overheating
Observe the permissible ambient conditions.
Ensure sufficient ventilation, especially on the ventilation slots on the left and right
of the device: There should be free space in the front, to the rear and sides of the
unit of at least 10 cm.
Do not block the fan inlet in the accessory compartment.
Keep heat sources away from the device.
Make sure that the power switch or power plug is always easily accessible.
Disconnect the device from the mains immediately if smoke is emitted from the
device.
WARNING
Device may fall over due to earthquake
In areas where there is a risk of earthquakes, secure the device to the device
handle with a short chain or a tension belt (designed for 1100 N).
Fig.7: Handle of the device
CAUTION
Risk of injury due to shortening of the corrugated hose
A corrugated hose connected to the inlet is shortened during evacuation.
To avoid endangering the stability of the device during evacuation and to prevent
injuries caused by a hose bursting, mount the corrugated hose without mechanical
tension.
Vent the corrugated hose before loosening.