95
4. From the computer’s keyboard, type “?” repeatedly until the
6700 controller determines the baud rate. Once the baud rate
is known, the 6700 returns an identification banner. This ban-
ner indicates that the connection is working properly. If you
have difficulty with the connection, try a slower baud rate.
5. Type DATA<CR> each time the data string is desired.
The serial data is in comma-separated-value format. Only the val-
ues that the sampler is set up to measure will be output in the data
string. The string is terminated with a <CR><LF>. To read the
data in the string, refer to
Table 11
.
I
DENTIFIER
P
ARAMETER
U
NITS
B?
1
B
OTTLE
NUMBER
AND
TIME
days since 1-Jan 1900
CR C
URRENT
DAY
S
RAIN
tips (rolls over every 256 tips)
CS
2
C
HECKSUM
unsigned long
DE D
ESCRIPTION
6700 Sampler
FL F
LOW
cubic meters per second
ID U
NIT
SPECIFIC
IDENTIFIER
10 digit unsigned long
LE L
EVEL
meters
MO M
ODEL
6700
PE P
ERCENT
percent of full scale
PH
P
H none
PR P
REVIOUS
DAY
S
RAIN
tips (rolls over every 256 tips)
RA R
AIN
tips (rolls over every 256 tips)
RTE R
EFRIGERATOR
T
EMPERATURE
degrees celsius
SS S
AMPLER
ENABLE
STATE
logical
TE T
EMPERATURE
degrees celsius
TI C
URRENT
TIME
days since 1-Jan-1900
VE V
ELOCITY
meters per second
VO V
OLUME
cubic meters
VSI V
ELOCITY
SIGNAL
STRENGTH
percent
VSP V
ELOCITY
SPECTRUM
STRENGTH
percent
YCO YSI 600 C
ONDUCTIVITY
milliseimens per centimeter
YDO YSI 600 D.O. milligrams per liter
YPH YSI 600
P
H none
YSA YSI 600 S
ALINITY
parts per thousand
YSP YSI 600 S
PECIFIC
C
ONDUCTANCE
milliseimens per centimeter
YTD YSI 600 T
OTAL
D
ISSOLVED
S
OLIDS
grams per liter
YTE YSI 600 T
EMPERATURE
degrees celsius
N
OTES
1. T
HE
THREE
MOST
RECENT
SAMPLES
WILL
HAVE
ENTRIES
AS
B?. T
HE
“?”
WILL
BE
THE
FIRST
BOTTLE
NUMBER
IN
THE
SET
.
IF
THE
SAMPLE
IS
A
GRAB
SAMPLE
,
IT
WILL
BE
SHOWN
AS
B0.
2. T
HE
CHECKSUM
DOES
NOT
INCLUDE
THE
CHECKSUM
,
CARRIAGE
RETURN
,
AND
LINEFEED
.
Connecting either the serial out-
put or the standard interrogator
cable to the 6700 disables the
optional dial-out modem, if in-
stalled. The interrogator cable
must be disconnected before
you can use the modem.
Note
Table 11 Serial Data Codes

Table of Contents