EasyManua.ls Logo

ISCO 6700SR - Resetting the Internal Battery Warning

ISCO 6700SR
190 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6700SR Refrigerated Sampler
76
1
RUN
PROGRAM
VIEW REPORT
OTHER FUNCTIONS
Select OTHER FUNCTIONS.
2
MAINTENANCE
MANUAL FUNCTIONS
PROGRAMMING STYLE
HARDWARE
Select MAINTENANCE.
3
SET CLOCK
SET PUMP ALARM
INTERNAL BATTERY
DIAGNOSTICS
Select INTERNAL BATTERY.
4
WARNING!
REPLACE INTERNAL
BATTERY BEFORE
17-OCT-97
The sampler displays this screen briefly to show you the expiration date.
5
DID YOU REPLACE THE
INTERNAL BATTERY?
YES NO
After replacing the battery, select YES. Select NO only if checking the
expiration date.
Resetting the alarm directs the sampler to calculate a new expiration date,
five years in the future. If you reset the alarm without replacing the battery,
the sampler will not be able to alert you to the battery’s impending
expiration.
6
MAINTENANCE
MANUAL FUNCTIONS
PROGRAMMING STYLE
HARDWARE
Press Stop to return to the main menu.
Example 8 Resetting the Internal Battery Warning

Table of Contents