6700SR Refrigerated Sampler
97
Memory
The 6700 Sampler contains 128 kilobytes of battery backed RAM
(Random Access Memory) that stores:
Five sampling programs
The sampling results from the most recent program
The rainfall data collected from rain gauges
Temperature data collected from the Refrigerator Temperature
Sensor
The data collected from a 700 Series module
Parameter data collected from a YSI Sonde
Calculated flow information from the current flow module
The sampling results memory stores up to 1004 events. The pro-
gram run and start time information is logged as four events. These
four entries are held for the duration of the program. The remain-
ing 1000 are for logging sample events, program events, and errors
from the currently running program. If the running program ex-
ceeds 1000 events, the data will “roll-over.” That is, the sampler
will replace the oldest stored event with the newest event. When
using a YSI Sonde the sampler logs two entries at each sample
event. This reduces the sampling results memory to 500 events.
Module, YSI Sonde, Refrigerator Temperature, and rain gauge
readings are stored in memory partitions. The storage capacity de-
pends on the reading types, the number of bytes per reading, and
the data storage interval.
The sampler clears the sampling results memory when the site de-
scription changes. It is important to collect the sampling results be-
fore changing the site description – the data can not be retrieved
afterwards. Keep in mind that the description may change when
selecting an extended program.
The partition data is cleared when the sampler automatically cre-
ates and sizes the memory partitions. This occurs when you:
Select a new program
Change the Data Storage Interval in the current program
Change the programming mode
Install a new 700 Series Module
Change 700 Series Module mode of operation
Add or remove a rain gauge with the Hardware Setup
Add or remove the Refrigerator Temperature Sensor as I/O3
Add or remove a YSI Sonde parameter with the Hardware Setup
Reinitialize the sampler
If you need to collect sampling results and partition data, do this
before allowing the sampler to create and size the partitions. The
stored data is lost as the sampler reformats its memory. As a pre-
caution, the sampler displays the screen below.
INTERVAL CHANGED --
STORED DATA WILL
BE LOST! OK?
YES
NO