6700SR Refrigerated Sampler
15
Some screens display the range of acceptable numbers between pa-
rentheses. If you enter a number that is too low or high, the control-
ler beeps and erases the entry. Type a new number and continue.
Stored Data Will Be Lost
Changes to the data storage interval, module setup, hardware set-
up, or selecting a new stored program, make it necessary for the
sampler to reformat its memory. See Memory on page 97. Refor-
matting the memory clears any stored data (sample events and
readings recorded during the last running program). As a precau-
tion, the sampler displays a message (screen d).
If you do not need the stored data, select YES. The sampler will re-
format its memory according to your menu selection and display
the new memory capacity.
If you need to collect or view the data, select NO. The sampler will
ignore your last menu selection(s) and retain the stored data. Exit
programming by pressing the STOP key and collect the data. See
Sampling Reports on page 54.
1
6700 SAMPLER
STANDARD PROGRAMMING
For HELP at any
screen press ? key.
Turn the sampler on by pressing the
On/Off
key. Press
↵
(Enter).
This screen disappears on its own after 8 seconds.
2
RUN
PROGRAM
VIEW REPORT
OTHER FUNCTIONS
The option PROGRAM will be blinking. Press
↵
(Enter).
3
SITE DESCRIPTION:
“FACTORY051“
CHANGE?
YES
NO
The option NO will be blinking. Press
↵
(Enter).
For the purposes of this example, it isn’t necessary to change the description.
4
SELECT UNITS FOR LENGTH:
ft
m
The option ft will be blinking, press
↵
(Enter).
5
DATA STORAGE
INTERVAL IN MINUTES
1
25
10 15 30
The “1” option will be blinking. Press
↵
(Enter).
6
NUMBER OF BOTTLES:
124812
24
Select the number of bottles in your bottle kit by pressing either arrow key
until the correct number blinks. Press
↵
(Enter). For this example, select 24.
7
BOTTLE VOLUME IS
1000 ml (300-30000)
Type the volume for the bottles in your kit. For this example, 1000 is correct,
so simply press
↵
(Enter).
Entering Numbers
c
SUCTION LINE LENGTH
IS __ ft
(3-99)
•Type the length of the suction
line. Acceptable values are
shown between the
parentheses. Press enter to
accept the number.
d
INTERVAL CHANGED --
STORED DATA WILL
BE LOST! OK?
YES
NO
Example 1 Program: One Sample every 15 Minutes, One Sample in Each Bottle Using
Normal Programming Style