6700SR Refrigerated Sampler
16
Example 1(Continued)
8
SUCTION LINE LENGTH
IS 10 ft
(3-99)
Type the length of the suction line, then press
↵
(Enter).
If you change the length, the sampler will display a message,
“PLEASE WAIT!. . . GENERATING PUMP TABLES.”
9
TIME PACED
FLOW PACED
Because this example requires samples every 15 minutes, select TIME
PACED by pressing an arrow until the option TIME PACED blinks. Then,
press
↵
(Enter).
10
TIME BETWEEN
SAMPLE EVENTS
0 HOURS, 15 MINUTES
Type 0 for hours and press
↵
(Enter). Type 15 for minutes and press
↵
(Enter).
Tip: Move back and forth between hours and minutes by pressing an arrow
key.
11
SEQUENTIAL
BOTTLES/SAMPLE
SAMPLES/BOTTLE
Because this program requires one sample in each bottle, select
SEQUENTIAL by pressing an arrow until the option SEQUENTIAL blinks.
Then, press
↵
(Enter).
12
RUN CONTINUOUSLY?
YES
NO
For this example, select NO by pressing an arrow until the option blinks.
Then, press
↵
(Enter). Selecting YES allows the program to run indefinitely
by repeating the sample distribution. Continuous sampling assumes that
filled bottles are replaced with empty bottles at regular service intervals.
13
SAMPLE VOLUME:
200 ml (10-1000)
Type the volume of the sample you want deposited in each bottle. Then,
press
↵
(Enter).
14
NO DELAY TO START
DELAYED START
CLOCK TIME
For this example, select DELAYED START by pressing an arrow until the
option blinks. Then, press
↵
(Enter).
15
FIRST SAMPLE
AFTER A
5 MINUTE DELAY
(1-999)
Type the delay period you want between the time you run the program and
the time the sampler takes the first sample. Then, press
↵
(Enter).
16
PROGRAMMING COMPLETE
RUN THIS PROGRAM
NOW?
YES
NO
Run the program immediately by selecting YES. Select NO if you want to
run the program later by selecting RUN from the main menu. Press
↵
(Enter) after making your choice. In this example, NO is selected.
17
RUN
PROGRAM
VIEW REPORT
OTHER FUNCTIONS
Run the program by selecting RUN and pressing
↵
(Enter).