6700SR Refrigerated Sampler
50
Samplers with an attached YSI Sonde will display the parameter
readings selected during the YSI hardware setup. The YSI screens
alternate with the sampler and module screens.
The Module and YSI readings are updated:
Every 5 seconds when not running a program and not display-
ing the reading.
Every 5 seconds when running a program and the backlight is
on.
Every 30 seconds when running a program and the reading is
used for pacing or enabling the sampler.
At the programmed storage interval when running a program
and the reading is not used for pacing or enabling the sampler.
Every second when displaying the reading during program-
ming.
An (asterisk) appears next to the reading if the module or YSI
Sonde was unable to take a reading. If an asterisk appears, the
reading displayed is the last available reading.
If an entire reading is filled with asterisks, the value exceeds the
number of characters that can be displayed. The sampler stores the
actual reading in memory, so it may be collected later.
For those programs that have delayed or scheduled start times, the
readings will be displayed while waiting for the start time. Keep in
mind that any totalizer values will be reset at the start time. Mod-
ule and YSI readings are also displayed after a program is com-
plete, until a key is pressed.
It is not necessary to run a program to view module or YSI read-
ings. Real time data can be viewed on the display by selecting
“VIEW REPORT” from the main menu. See Viewing the Data on
page 54.
Error Messages
If the sampler detects an error that prevents it from taking a sam-
ple or continuing the program, it displays an error message.
There are messages for the following errors:
Pump Jammed No Liquid Detected
Distributor Jammed No More Liquid
Probable Overflow Pump Latch open
Power Failed Sampler Shut Off
User Stopped Bottle Full
No Distributor Arm Sample In Progress
BOTTLE 2
IN 00:14:39
150 gpm 0.82 ft
00001438200 gal
Y-pH: 8.1
Y-D.O: 4.90 mg/l
Y-TDS: 1.000 g/l
Y-TEMP: 29.2 C
Error Messages
i
BOTTLE 2
IN 00:12:00
❋❋❋❋❋
gpm 5.73 ft
00001457600 gal
j
ERRORS HAVE OCCURED
DURING PROGRAM
k
FATAL ERROR
NO DISTRIBUTOR ARM

Table of Contents