EasyManua.ls Logo

ISCO 6700SR - Page 45

ISCO 6700SR
190 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6700SR Refrigerated Sampler
27
Example 5 Standard Program: Flow-Paced Sampling, Two
Bottles Per Sample
Program type: Standard
Site description: SITE 29
Units for Length: Feet
Data Storage Interval: 15 minutes
Bottle kit: 24, 1000 milliliter bottles
Suction line length: 7 feet
Pacing: Flow pacing, every two pulses
Distribution: 2 bottles per sample
Sample Volume: 250 milliliters
Start time: Clock Time, 6:00 a.m. on Monday, Wednesday, Fri-
day
Module
:
No module installed
1
6700 SAMPLER
STANDARD PROGRAMMING
For HELP at any
screen press ? key.
2
RUN
PROGRAM
VIEW REPORT
OTHER FUNCTIONS
Select PROGRAM.
3
SITE DESCRIPTION
“SITE 29
CHANGE
YES
NO
Select NO.
(The margin example, Using the Site Description or Program Name Menus
on page 18, shows you how to change the site description.)
4
SELECT UNITS FOR
LENGTH
ft
m
Select ft.
5
DATA STORAGE
INTERVAL IN MINUTES
1 2 5
10
15
30
Select 15.
6
NUMBER OF BOTTLES
124812
24
Select 24.
7
BOTTLE VOLUME IS
1000
ml (300-30000)
Enter 1000.
8
SUCTION LINE LENGTH
IS
7
ft
(3-99)
Enter 7.
9
TIME PACED
FLOW PACED
Select FLOW PACED.

Table of Contents