Calibration Command
Copyright© Itech Electronics Co., Ltd. 97
Chapter14 Calibration Command
CALibrate:SECure[:STATe]
This command is used to set the calibration state.
Command syntax
CALibrate:SECure[:STATe] <bool>,<string>
Arguments
0|1|OFF|ON,<string>
Example
CAL:SEC 0, 6522
Query syntax
CALibrate:SECure[:STATe]?
Returns
0|1
CALibrate:VOLTage:LEVel
This command is used to define the voltage calibration point.
Command syntax
CALibrate:VOLTage:LEVel <P0|P1>
Arguments
P0|P1
Example
CAL:VOLT:LEV P0
CALibrate:VOLTage[:DATA]
This command is used to read the actual voltage of the present
calibration point.
Command syntax
CALibrate:VOLTage[:DATA] <NRf>
Arguments
NRf
Unit
V