Car Wave CommandIT6500C
Copyright© Itech Electronics Co., Ltd. 78
Returns
NR3
CARWave:ISO16750:LOAD:DUMP[:TIMe]:TD[:LEVel]
This command is used to set pulse width of load dump.
Command syntax
CARWave:ISO16750:LOAD:DUMP:TD <value>
Arguments
MINimum|MAXimum|NR3
Reset Value
0.004
Example
CARW:ISO16750:LOAD:DUMP:TD 40
Query syntax
CARW:ISO16750:LOAD:DUMP:TD? [MINimum|MAXimum]
Returns
NR3
CARWave:ISO16750:LOAD:DUMP:VOLTage[:LEVel]
This command is used to set the voltage system of load dump.
Command syntax
CARWave:ISO16750:LOAD:DUMP:VOLT <12V|24V>
Arguments
12V|24V
Reset Value
12V
Example
CARW:ISO16750:LOAD:DUMP:VOLT 12V
Query syntax
CARW:ISO16750:LOAD:DUMP:VOLT?
Returns
<CRD>