Output Commands
Copyright© Itech Electronics Co., Ltd. 36
[SOURce:]VOLTage:PROTection:STATe
This command is used to set over voltage protection state.
Command syntax
[SOURce:]VOLTage:PROTection:STATe <bool>
Arguments
ON|OFF|1|0
Query syntax
[SOURce:]VOLTage:PROTection:STATe?
Returns
0|1
[SOURce:]VOLTage:PROTection[:LEVel]
This command is used to set voltage value of OVP.
Command syntax
[SOURce:]VOLTage:PROTection[:LEVel] <MINimum|MAXimum|NRf>
Arguments
MINimum|MAXimum|NRf
Unit
V
Reset Value
MAXimum
Example
VOLT:PROT MAX
Query syntax
[SOURce:]VOLTage:PROTection[:LEVel]? [MINimum|MAXimum]
Returns
NRF
[SOURce:]VOLTage:PROTection:DELay
This command is used to set delay time for over voltage protection.
Command syntax
[SOURce:]VOLTage:PROTection:DELay <MINimum|MAXimum|NRf>