EasyManua.ls Logo

ITech IT6500C Series

ITech IT6500C Series
126 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Calibration Command
Copyright© Itech Electronics Co., Ltd. 98
Example
CAL:VOLT 12.0V
CALibrate:CURRent:LEVel
This command is used to define the current calibration points.
Command syntax
CALibrate:CURRent:LEVel <P0|P1>
Arguments
P0|P1
Example
CAL:CURR:LEV P0
CALibrate:CURRent[:DATA]
This command is used to read the actual current of the present
calibration point.
Command syntax
CALibrate:CURRent[:DATA] <NRf>
Arguments
NRf
Unit
A
Example
CAL:CURR 20.0
CALibrate:MONitor:VOLTage:LEVel
This command is used to set monitoring the voltage calibration point.
Command syntax
CALibrate:MONitor:VOLTage:LEVel <P0|P1>
Arguments
P0|P1
Example
CAL:MON:VOLT:LEV P0
CALibrate:MONitor:VOLTage[:DATA]
This command is used to set the voltage calibration point data for monitoring.

Table of Contents

Other manuals for ITech IT6500C Series

Related product manuals