EasyManuals Logo

Keithley 2015 Service Manual

Keithley 2015
158 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #120 background imageLoading...
Page #120 background image
A-6 Specifications
TRUE RMS AC VOLTAGE AND CURRENT CHARACTERISTICS
ACCURACY
1
: ±(% of reading + % of range), 23°C ±5 °C
VOLTAGE CALIBRATION 3 Hz– 10 Hz– 20 kHz– 50 kHz– 100 kHz–
RANGE RESOLUTION CYCLE 10 Hz 20 kHz 50 kHz 100 kHz 300 kHz
100.0000 mV 0.1 µV
1.000000 V 1.0 µV 90 Days 0.35 + 0.03 0.05 + 0.03 0.11 + 0.05 0.60 + 0.08 4 + 0.5
10.00000 V 10 µV
100.0000 V 100 µV 1 Year 0.35 + 0.03 0.06 + 0.03 0.12 + 0.05 0.60 + 0.08 4 + 0.5
750.000 V 1 mV
TEMPERATURE
COEFFICIENT
8
0.035 + 0.003 0.005 + 0.003 0.006 + 0.005 0.01 + 0.006 0.03 + 0.01
CURRENT CALIBRATION 3 Hz - 10 Hz -
RANGE RESOLUTION CYCLE 10 Hz 5 kHz
1.000000 A 1 µA 90 Day/1 Year 0.30 + 0.04 0.10 + 0.04
3.00000 A 10 µA 90 Day/1 Year 0.35 + 0.06 0.15 + 0.06
TEMPERATURE
COEFFICIENT
8
0.035 + 0.006 0.015 + 0.006
HIGH CREST FACTOR ADDITIONAL ERROR ±(% of reading)
7
CREST FACTOR: 1–2 2–3 3–4 4–5
ADDITIONAL ERROR: 0.05 0.15 0.30 0.40
AC OPERATING CHARACTERISTICS
2
FUNCTION DIGITS READINGS/s RATE BANDWIDTH
ACV (all ranges), and 6
1
2
3
2s/reading SLOW 3 Hz–300 kHz
ACI (all ranges) 6
1
2
3
1.4 MED 30 Hz–300 kHz
6
1
2
4
4.8 MED 30 Hz–300 kHz
6
1
2
3
2.2 FAST 300 Hz–300 kHz
6
1
2
4
35 FAST 300 Hz–300 kHz
ADDITIONAL LOW FREQUENCY ERRORS ±(% of reading)
SLOW MED FAST
20Hz – 30Hz 0 0.3
30Hz – 50Hz 0 0
50Hz – 100Hz 0 0 1.0
100Hz – 200Hz 0 0 0.18
200Hz – 300Hz 0 0 0.10
> 300Hz 0 0 0
AC SYSTEM SPEEDS
2, 5
FUNCTION/RANGE CHANGE
6
: 4 / s.
AUTORANGE TIME: <3 s.
ASCII READINGS TO RS-232 (19.2K BAUD)
4
: 50 / s.
MAX. INTERNAL TRIGGER RATE
4
: 300 / s.
MAX. EXTERNAL TRIGGER RATE
4
: 260 / s.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2015 and is the answer not in the manual?

Keithley 2015 Specifications

General IconGeneral
ModelKeithley 2015
CategoryMultimeter
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Frequency Range3 Hz to 300 kHz
Resistance Range100 MΩ
DC Current Range3 A
DC Current Accuracy0.05%
AC Current Range3 A
Display5.5 digit
ConnectivityGPIB
Power Supply100 V to 240 V AC, 50 Hz to 60 Hz