Calibration Reference B-7
:LOCK?
(:CALibration:PROTected:LOCK?)
Purpose
To read comprehensive calibration lock status.
Format
:cal:prot:lock?
Response
0 Comprehensive calibration locked.
1 Comprehensive calibration unlocked.
Description
The
:LOCK?
query requests status from the Model 2015 on calibration
locked/unlocked state. Calibration must be enabled sending the
:CODE
com-
mand before calibration can be performed.
Example
:CAL:PROT:LOCK?
Request cal lock state.
:SAVE
(:CALibration:PROTected:SAVE)
Purpose
To save calibration constants in EEROM after the calibration procedure.
Format
:cal:prot:save
Parameter
None
Description
The
:SAVE
command stores internally calculated calibration constants
derived during both comprehensive and manufacturing calibration in
EEROM. (EEROM is non-volatile memory.) Calibration constants will be
retained indefinitely once saved. Generally,
:SAVE
is sent after all other cal-
ibration steps (except for
:LOCK
).
NOTE
Calibration will be only temporary unless the
:SAVE
command is sent to permanently
store calibration constants.
Example
:CAL:PROT:SAVE
Save calibration constants.