Calibration 2-23
Function generator calibration
1. Connect the rear panel SOURCE OUTPUT jack to the front panel INPUT jacks (see Fig-
ure 2-6.)
2. Send the following command to complete function generator calibration:
:CAL:PROT:FGEN:STEP1
Programming calibration dates
Program the present calibration date and calibration due date by sending the following
commands:
:CAL:PROT:DATE <year>, <month>, <day>
:CAL:PROT:NDUE <year>, <month>, <day>
For example, the following commands assume calibration dates of 12/15/97 and 3/14/98
respectively:
:CAL:PROT:DATE 1997, 12, 15
:CAL:PROT:NDUE 1998, 3, 14
Saving calibration constants
After completing the calibration procedure, send the following command to save the new cal-
ibration constants:
:CAL:PROT:SAVE
NOTE Calibration constants will not be saved unless the :SAVE command is sent.
Locking out calibration
After saving calibration, send the following command to lock out calibration:
:CAL:PROT:LOCK

Table of Contents

Question and Answer IconNeed help?

Do you have a question about the Keithley 2015 and is the answer not in the manual?

Keithley 2015 Specifications

General IconGeneral
ModelKeithley 2015
CategoryMultimeter
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Frequency Range3 Hz to 300 kHz
Resistance Range100 MΩ
DC Current Range3 A
DC Current Accuracy0.05%
AC Current Range3 A
Display5.5 digit
ConnectivityGPIB
Power Supply100 V to 240 V AC, 50 Hz to 60 Hz

Summary

WARRANTY

Safety Precautions

Performance Verification

Verifying DC voltage

Describes the procedure to check DC voltage accuracy using a calibrator.

Verifying AC voltage

Details the process for verifying AC voltage accuracy at specific frequencies.

Verifying resistance

Details the procedure for checking resistance accuracy using a calibrator.

Verifying frequency

Explains the steps to verify the Model 2015's frequency measurement function.

Verifying total harmonic distortion

Details the procedure to verify the total harmonic distortion measurement.

Calibration

Calibration code

Explains how to access and enter the calibration code for unlocking.

Comprehensive calibration

Describes the process for calibrating multiple functions of the Model 2015.

DC volts calibration

Outlines the procedure for calibrating the DC volts measurement function.

Resistance calibration

Explains the steps to calibrate the resistance measurement function.

AC voltage calibration

Describes how to calibrate the AC voltage measurement function.

Routine Maintenance

Setting the line voltage and replacing the line fuse

Details how to set line voltage and replace the main power line fuse.

Replacing the AMPS fuse

Explains the procedure for replacing the AMPS fuse for over-current protection.

Troubleshooting

Repair considerations

Provides important advice and precautions before undertaking any repairs.

Disassembly

Case cover removal

Explains how to remove the instrument's outer case cover.

Specifications

Calibration Reference