1-8 Performance Verification
Verifying DC voltage
Check DC voltage accuracy by applying accurate voltages from the DC voltage calibrator to
the Model 2015 INPUT jacks and verifying that the displayed readings fall within specified
limits.
CAUTION
Do not exceed 1100V peak between INPUT HI and INPUT LO because
instrument damage may occur.
Follow these steps to verify DC voltage accuracy:
1. Connect the Model 2015 HI and LO INPUT jacks to the DC voltage calibrator as shown
in Figure 1-1.
NOTE
Use shielded, low-thermal connections when testing the 100mV and 1V ranges to
avoid errors caused by noise or thermal effects. Connect the shield to the calibrator’s
output LO terminal.
Figure 1-1
Connections for DC volts verification
Output HI
Output
LO
2015 THD MULTIMETER
RANGE
!
F
500V
PEAK
FRONT/REAR
3A 250V
AMPS
HI
INPUT
LO
SENSE
4 WIRE
INPUTS
350V
PEAK
1000V
PEAK
AUTO
SHIFT
LOCAL
POWER
RANGE
R
SHIFT
CH1REM
TALK
LSTN
SRQ
STAT
REL FILT
4W
BUFFER
MATH
REAR
SCAN
TIMER
STEP CH2 CH3 CH4 CH5 CH6 CH7 CH8 CH9 CH10
HOLD TRIG FAST MED SLOW AUTO ERR
EXIT ENTER
DIGITS RATE
RELFILTER
TRIG
EX TRIG
STORE
RECALL
SOURCE
MEAS
DCV
DCI
MATH
THD
dBm
ACV
ACI
2 4
FREQ
TEMP
dB
CONT
PERIOD TCOUPL
LIMITS ON/OFFDELAY
HOLD
SAVE SETUP
CONFIG HALT
TEST
RS232
GPIB
CAL
STEP SCAN
Model 2015
Input
LO
Input HI
DC Voltage Calibrator
Note: Use shielded, low-thermal cables
for 100mV and 1V ranges.
THD

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2015 and is the answer not in the manual?

Keithley 2015 Specifications

General IconGeneral
BrandKeithley
Model2015
CategoryMultimeter
LanguageEnglish

Related product manuals