Calibration Reference B-15
:AC:STEP<n>
(CALibration:PROTected:AC:STEP<n>)
Purpose To program individual AC calibration steps.
Format :cal:prot:ac:step<n>
Parameters 1 10mV AC at 1kHz calibration step.
2 100mV AC at 1kHZ calibration step.
3 100mV AC at 50kHz calibration step.
4 1V AC at 1kHz calibration step.
5 1V AC at 50kHz calibration step.
6 10V AC at 1kHz calibration step.
7 10V AC at 50kHz calibration step.
8 100V AC at 1kHz calibration step.
9 100V AC at 50khz calibration step.
10 700V AC at 1kHz calibration step.
11 100mA AC at 1kHz calibration step.
12 1A AC at 1kHz calibration step.
13 2A AC at 1kHz calibration step.
Description The :AC:STEP command programs the 13 individual AC calibration steps;
<n> represents the calibration step number. The appropriate signal must be
connected to the instrument when programming each step, as summarized in
the parameters listed above.
Example :CAL:PROT:AC:STEP7 Program 10V, 50kHz step.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2015 and is the answer not in the manual?

Keithley 2015 Specifications

General IconGeneral
ModelKeithley 2015
CategoryMultimeter
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Frequency Range3 Hz to 300 kHz
Resistance Range100 MΩ
DC Current Range3 A
DC Current Accuracy0.05%
AC Current Range3 A
Display5.5 digit
ConnectivityGPIB
Power Supply100 V to 240 V AC, 50 Hz to 60 Hz

Related product manuals