EasyManuals Logo
Home>Keithley>Multimeter>2015

Keithley 2015 Service Manual

Keithley 2015
158 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #138 background imageLoading...
Page #138 background image
B-12 Calibration Reference
:STEP8
(:CALibration:PROTected:DC:STEP8)
Purpose To program the 100k 4-wire comprehensive calibration step.
Format :cal:prot:dc:step8 <Cal_resistance>
Parameter <Cal_resistance> = 90E3 to 110E3 []
Description :STEP8 programs the 100k 4-wire resistance comprehensive calibration
step. The allowable range of the calibration resistance parameter is from
90E3 to 110E3, but 100E3 is recommended for best results.
Example :CAL:PROT:DC:STEP8 100E3 Program 100k step.
:STEP9
(CALibration:PROTected:DC:STEP9)
Purpose To program the 1M comprehensive calibration step.
Format :cal:prot:dc:step9 <Cal_resistance>
Parameter <Cal_resistance> = 900E3 to 1.1E6 []
Description :STEP9 programs the 1M comprehensive calibration step. The allowable
range of the calibration resistance parameter is from 900E3 to 1.1E6. Use the
1E6 value whenever possible, or the closest possible value.
Example :CAL:PROT:DC:STEP9 Program 1M calibration step.
:STEP10
(CALibration:PROTected:DC:STEP10)
Purpose To program the 10mA comprehensive calibration step.
Format :cal:prot:dc:step10 <Cal_current>
Parameter <Cal_current> = 9E-3 to 11E-3 [A]
Description :STEP10 programs the 10mA comprehensive calibration step. The allow-
able range of the calibration current parameter is from 9E-3 to 11E-3. Use
the 10E-3 value whenever possible for best results.
Example :CAL:PROT:DC:STEP10 10E-3 Program 10mA step.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2015 and is the answer not in the manual?

Keithley 2015 Specifications

General IconGeneral
BrandKeithley
Model2015
CategoryMultimeter
LanguageEnglish

Related product manuals