EasyManuals Logo

Keithley 2015 Service Manual

Keithley 2015
158 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #139 background imageLoading...
Page #139 background image
Calibration Reference B-13
:STEP11
(CALibration:PROTected:DC:STEP11)
Purpose To program the 100mA comprehensive calibration step.
Format :cal:prot:dc:step11 <Cal_current>
Parameter <Cal_current> = 90E-3 to 110E-3 [A]
Description :STEP11 programs the 100mA comprehensive calibration step. The allow-
able range of the calibration current parameter is from 90E-3 to 110E-3. Use
the 100E-3 value whenever possible for best results.
Example :CAL:PROT:DC:STEP11 0.1 Program 100mA step.
:STEP12
(CALibration:PROTected:DC:STEP12)
Purpose To program the 1A comprehensive calibration step.
Format :cal:prot:dc:step12 <Cal_current>
Parameter <Cal_current> = 0.9 to 1.1 [A]
Description :STEP12 programs the 1A comprehensive calibration step. The allowable
range of the calibration current parameter is from 0.9 to 1.1. Use a value of
1 whenever possible for best results.
Example :CAL:PROT:DC:STEP12 1 Program 1A step.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2015 and is the answer not in the manual?

Keithley 2015 Specifications

General IconGeneral
ModelKeithley 2015
CategoryMultimeter
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Frequency Range3 Hz to 300 kHz
Resistance Range100 MΩ
DC Current Range3 A
DC Current Accuracy0.05%
AC Current Range3 A
Display5.5 digit
ConnectivityGPIB
Power Supply100 V to 240 V AC, 50 Hz to 60 Hz

Related product manuals