Specifications A-9
Accuracy calculations
The information below discusses how to calculate accuracy for both DC and AC characteristics.
Calculating DC characteristics accuracy
DC characteristics accuracy is calculated as follows:
Accuracy = ±(ppm of reading + ppm of range)
(ppm = parts per million, and 10ppm = 0.001%)
As an example of how to calculate the actual reading limits, assume that you are measuring
5V on the 10V range. You can compute the reading limit range from one-year DCV accuracy
specifications as follows:
Accuracy = ±(30ppm of reading + 5ppm of range)
±[(30ppm
×
5V) + (5ppm
×
10V)]
±(150µV + 50µV)
±200µV
Thus, the actual reading range is: 5V± 200µV, or from 4.9998V to 5.0002V
DC current and resistance calculations are performed in exactly the same manner using the
pertinent specifications, ranges, and input signal values.
Calculating AC characteristics accuracy
AC characteristics accuracy is calculated similarly, except that AC specifications are given as
follows:
Accuracy = ±(% of reading + % of range)
As an example of how to calculate the actual reading limits, assume that you are measuring
120V, 60Hz on the 750V range. You can compute the reading limit range from ACV one-year
accuracy specifications as follows:
Accuracy = ±(0.06% of reading + 0.03% of range)
±[(0.0006
×
120V) + (0.0003
×
750V)]
±(0.072V + 0.225V)
±0.297V
In this case, the actual reading range is: 120V ± 0.297V, or from 119.703V to 120.297V
AC current calculations are performed in exactly the same manner using the pertinent speci-
fications, ranges, and input signal values.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2015 and is the answer not in the manual?

Keithley 2015 Specifications

General IconGeneral
ModelKeithley 2015
CategoryMultimeter
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Frequency Range3 Hz to 300 kHz
Resistance Range100 MΩ
DC Current Range3 A
DC Current Accuracy0.05%
AC Current Range3 A
Display5.5 digit
ConnectivityGPIB
Power Supply100 V to 240 V AC, 50 Hz to 60 Hz