Calibration Reference B-5
Miscellaneous calibration commands
Miscellaneous commands perform miscellaneous calibration functions such as programming
the calibration code and date. These commands are discussed in detail in the following para-
graphs.
:CODE
(:CALibration:PROTected:CODE)
Purpose
To program the calibration code or password so that you can perform the cal-
ibration procedures.
Format
:cal:prot:code '<char_string>'
Parameter
Up to a 8-character string including letters and numbers.
Description
The
:CODE
command enables the Model 2015 calibration procedures when
performing these procedures over the bus. In general, this command must be
sent to the unit before sending any other comprehensive or manufacturing
calibration command. The default calibration code is KI002015.
NOTES
The
:CODE
command should be sent only once before performing either the compre-
hensive or factory calibration. Do not send
:CODE
before each calibration step.
To change the code, first send the present code, then send the new code.
The code parameter must be enclosed in single quotes.
Example
:CAL:PROT:CODE 'KI002015'
Send default code of KI002015.
:COUNt?
(:CALibration:PROTected:COUNt?)
Purpose
To determine how many times the Model 2015 has been calibrated.
Format
:cal:prot:coun?
Response
<n> Calibration count.
Description
The
:COUNt?
command allows you to determine how many times the
Model 2015 has been calibrated.
NOTE
Use the
:COUNt?
command to help you monitor for unauthorized calibration
procedures.
Example
:CAL:PROT:COUN?
Request calibration count.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2015 and is the answer not in the manual?

Keithley 2015 Specifications

General IconGeneral
ModelKeithley 2015
CategoryMultimeter
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Frequency Range3 Hz to 300 kHz
Resistance Range100 MΩ
DC Current Range3 A
DC Current Accuracy0.05%
AC Current Range3 A
Display5.5 digit
ConnectivityGPIB
Power Supply100 V to 240 V AC, 50 Hz to 60 Hz

Related product manuals