EasyManuals Logo

Keithley 2015 Service Manual

Keithley 2015
158 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #130 background imageLoading...
Page #130 background image
B-4 Calibration Reference
Table B-1
Remote calibration command summary (continued)
Command Description
:CALibration Calibration root command.
:PROTected
:AC AC cal steps.
:STEP1 10mV AC at 1kHz step.
:STEP2 100mV AC at 1kHz step.
:STEP3 100mV AC at 50kHz step.
:STEP4 1V AC at 1kHz step.
:STEP5 1V AC at 50kHz step.
:STEP6 10V AC at 1kHz step.
:STEP7 10V AC at 50kHz step.
:STEP8 100V AC at 1kHz step.
:STEP9 100V AC at 50kHz step.
:STEP10 700V AC at 1kHz step.
:STEP11 100mA AC at 1kHz step.
:STEP12 1A AC at 1kHz step.
:STEP13 2A AC at 1kHz step
1
.
:STEP14 1V AC at 3Hz step
1
.
:STEP15 1V AC at 1kHz step
1
.
:DIST Distortion cal steps.
:STEP1 1V RMS at 137Hz step.
:STEP2 1V RMS at 844Hz step.
:FGEN
:STEP1 Calibrate function generator.
NOTES:
1. DC:STEP0, AC:STEP14, and AC:STEP15 are one-time factory calibration points and are valid only in man-
ufacturing calibration mode.
2. Upper-case letters indicated short form of each command. For example, instead of sending
“:CALibration:PROTected:INITiate,” you can send “:CAL:PROT:INIT.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2015 and is the answer not in the manual?

Keithley 2015 Specifications

General IconGeneral
ModelKeithley 2015
CategoryMultimeter
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Frequency Range3 Hz to 300 kHz
Resistance Range100 MΩ
DC Current Range3 A
DC Current Accuracy0.05%
AC Current Range3 A
Display5.5 digit
ConnectivityGPIB
Power Supply100 V to 240 V AC, 50 Hz to 60 Hz

Related product manuals