B-4 Calibration Reference
Table B-1
Remote calibration command summary (continued)
Command Description
:CALibration Calibration root command.
:PROTected
:AC AC cal steps.
:STEP1 10mV AC at 1kHz step.
:STEP2 100mV AC at 1kHz step.
:STEP3 100mV AC at 50kHz step.
:STEP4 1V AC at 1kHz step.
:STEP5 1V AC at 50kHz step.
:STEP6 10V AC at 1kHz step.
:STEP7 10V AC at 50kHz step.
:STEP8 100V AC at 1kHz step.
:STEP9 100V AC at 50kHz step.
:STEP10 700V AC at 1kHz step.
:STEP11 100mA AC at 1kHz step.
:STEP12 1A AC at 1kHz step.
:STEP13 2A AC at 1kHz step
1
.
:STEP14 1V AC at 3Hz step
1
.
:STEP15 1V AC at 1kHz step
1
.
:DIST Distortion cal steps.
:STEP1 1V RMS at 137Hz step.
:STEP2 1V RMS at 844Hz step.
:FGEN
:STEP1 Calibrate function generator.
NOTES:
1. DC:STEP0, AC:STEP14, and AC:STEP15 are one-time factory calibration points and are valid only in man-
ufacturing calibration mode.
2. Upper-case letters indicated short form of each command. For example, instead of sending
“:CALibration:PROTected:INITiate,” you can send “:CAL:PROT:INIT.”