Calibration 2-5
Calibration code
Before performing comprehensive calibration, you must first unlock calibration by entering
the appropriate calibration code.
Front panel calibration code
For front panel calibration, follow these steps:
1. Access the calibration menu by pressing SHIFT CAL, and note that the instrument dis-
plays the following:
CAL: DATES
2. Use the up or down range key to scroll through the available calibration parameters until
the unit displays RUN, then press ENTER.
3. The Model 2015 then prompts you to enter a code:
CODE? 000000
(The factory default code is 002015.) Use the left and right arrow keys to move among
the digits; use the up range key to increment numbers, and press the down range key to
specify alphabetic letters. Confirm the code by pressing ENTER.
4. The Model 2015 allows you to define a new calibration code. Use the up and down range
keys to toggle between yes and no. Choose N if you do not want to change the code.
Choose Y if you want to change the code. The unit then prompts you to enter a new code.
Enter the code, and press ENTER.
Remote calibration code
If you are performing calibration over the IEEE-488 bus or the RS-232 port, send this com-
mand to unlock calibration:
:CAL:PROT:CODE '<8-character string>'.
The default code command is:
:CAL:PROT:CODE 'KI002015'.