Page #66 background imageLoading...
Page #66 background image
Troubleshooting 4-3
Repair considerations
Before making any repairs to the Model 2015, be sure to read the following considerations.
CAUTION
The PC-boards are built using surface mount techniques and require spe-
cialized equipment and skills for repair. If you are not equipped and/or
qualified, it is strongly recommended that you send the unit back to the fac-
tory for repairs or limit repairs to the PC-board replacement level. Without
proper equipment and training, you could damage a PC-board beyond
repair.
Repairs will require various degrees of disassembly. However, it is recommended that
the Front Panel Tests be performed prior to any disassembly. The disassembly instruc-
tions for the Model 2015 are contained in
Section 5
of this manual.
Do not make repairs to surface mount PC-boards unless equipped and qualified to do so
(see previous CAUTION).
When working inside the unit and replacing parts, be sure to adhere to the handling pre-
cautions and cleaning procedures explained in
Section 5
.
Many CMOS devices are installed in the Model 2015. These static-sensitive devices
require special handling as explained in
Section 5
.
Whenever a circuit board is removed or a component is replaced, the Model 2015 must
be recalibrated. See
Section 2
for details on calibrating the unit.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2015 and is the answer not in the manual?

Keithley 2015 Specifications

General IconGeneral
ModelKeithley 2015
CategoryMultimeter
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Frequency Range3 Hz to 300 kHz
Resistance Range100 MΩ
DC Current Range3 A
DC Current Accuracy0.05%
AC Current Range3 A
Display5.5 digit
ConnectivityGPIB
Power Supply100 V to 240 V AC, 50 Hz to 60 Hz