Disassembly 5-11
Main CPU firmware replacement
Changing the firmware may be necessary as upgrades become available. The firmware revi-
sion levels for the main and front panel CPUs are displayed during the power-on sequence. (The
main firmware revision level is displayed on the left; the front panel firmware revision level is
displayed on the right.) For example: REV: A01 A02 indicates a main firmware revision level
of A01 and a front panel firmware revision level of A02.
The firmware for the main CPU is located in the EPROMs U156 (EVEN) and U157 (ODD),
leadless ICs that resides in chip carriers on the PC board. To replace the CPU firmware, do the
following:
WARNING
Disconnect the instrument from the power lines, and remove the test leads
before changing the firmware.
1. Remove the case cover as described earlier in this section.
2. Locate U156 EVEN and U157 ODD (EPROMs) on the DMM (106) board. They are the
only devices installed in chip carriers (sockets).
CAUTION
EPROMs U156 and U157 are static-sensitive devices. Be sure to follow the
handling precautions explained in Static sensitive devices.
3. Using an appropriate chip extractor, remove U156 from its chip carrier.
4. Position the new U156 EPROM on the appropriate chip carrier. Make sure the notched
corner of the chip is aligned with the notch in the chip carrier.
NOTE
Be sure to install the correct EPROMs at the ODD and EVEN locations. The instru-
ment will not function if the EPROMs are installed in the wrong sockets.
5. With the EPROM properly positioned, push down on the chip until it completely seats
into the chip carrier.
6. Repeat steps 3 through 5 for EPROM U157.
7. After installation, make sure the instrument powers up normally before replacing the
cover.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2015 and is the answer not in the manual?

Keithley 2015 Specifications

General IconGeneral
ModelKeithley 2015
CategoryMultimeter
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Frequency Range3 Hz to 300 kHz
Resistance Range100 MΩ
DC Current Range3 A
DC Current Accuracy0.05%
AC Current Range3 A
Display5.5 digit
ConnectivityGPIB
Power Supply100 V to 240 V AC, 50 Hz to 60 Hz

Related product manuals